A recent campaign at the Palomar Observatory used a Traveling Collimated Beam Projector (TCBP) to measure the transmission curves of the ZTF telescope’s g, r, and i filters with unprecedented precision. The TCBP delivers a monochromatic parallel beam monitored by a photodiode to 0.01 % and a spectrograph to 0.1 nm. This instrument was used to scan each filter on all 16 CCDs, revealing variations of up to 3 nm across the focal plane and out‑of‑band leaks of about 0.5 % below 400 nm in the g band that were detected with a statistical significance exceeding 20σ.
The measured curves showed strong departures from the publicly available model, especially in the core structures and edge positions of the passbands. Cross‑comparison with the Supernova Legacy Survey (SNLS) filters, which were calibrated with the lower‑resolution SnDICE instrument, confirmed a systematic difference at the few‑per‑mille level in all three bands.
These findings illustrate the importance of high‑resolution, in‑situ calibration for wide‑field surveys. They also highlight the limited resolution of the SNLS calibrator, which uses only 24 sampling points in the visible range.
Figure 1(b) displays the mean wavelength per CCD for each ZTF band, showing the sub‑nanometre uncertainties achieved and the peak‑to‑peak variations across the focal plane that underpin the color offsets observed in reference stars.